mailRead multiple R1 rate files for different offsets frequencies.


Others Months | Index by Date | Thread Index
>>   [Date Prev] [Date Next] [Thread Prev] [Thread Next]

Header


Content

Posted by Troels Emtekær Linnet on October 22, 2013 - 21:51:

Hi Edward.

At the moment, the reading of R1 rate files are associated on the spectrometer frequency level.
As can be seen in:
sample_scripts/relax_disp/R1rho_analysis.py

> -----------------
> # Load the R1 data.
> relax_data.read(ri_id='500MHz', ri_type='R1', frq=500e6, file='R1_500MHz.out', dir=data_path, mol_name_col=1, res_num_col=2, res_name_col=3, spin_num_col=4, spin_name_col=5, data_col=6, error_col=7)
> ------------------

I have created several R1 rate files in:
test_suite/shared_data/dispersion/Kjaergaard_et_al_2013

If one do:
> --------------
> ls -v -d -1 rx/*/rx.out > rxfiles.txt
> set RXS=`cat rxfiles.txt`
> set CWD=`pwd`
> echo $CWD
>
> foreach I (`seq 1 ${#RXS}`)
> set RX=${RXS[$I]}; set DIRN=`dirname $RX`
> cd $DIRN
> set LINE=`grep "28         Q" rx.out`
> echo "$DIRN $LINE"
> cd $CWD
> end
> ----------------
I get:

> rx/0_35 None 28 Q 20 N 11.8294310864515 0.0977051738570798
> rx/0_39 None 28 Q 20 N 12.1872073441636 0.194000815320065
> rx/0_41 None 28 Q 20 N 11.9893516334575 0.163599788621843
> rx/0_43 None 28 Q 20 N 11.8263428708274 0.123690768122611
> rx/0_46 None 28 Q 20 N 11.9323504806714 0.0444745428316398
> rx/0_48 None 28 Q 20 N 11.8632811958325 0.114755798629533
> rx/500_46 None 28 Q 20 N 11.0724074058591 0.1538326259565
> rx/1000_41 None 28 Q 20 N 6.33513730898357 0.0336559336993769
> rx/1000_46 None 28 Q 20 N 8.77127968755774 0.0641417105624404
> rx/1000_48 None 28 Q 20 N 9.70375281785425 0.0172589388319522
> rx/2000_41 None 28 Q 20 N 3.108869297874 0.0496833972732356
> rx/2000_46 None 28 Q 20 N 4.9044902550994 0.0439321013896472
> rx/5000_46 None 28 Q 20 N 2.25502543028924 0.0194300877425236
> rx/10000_46 None 28 Q 20 N 1.93165081138783 0.00987269881003522

This shows that the R1 rate is quite different for offsets frequencies.

How should this problem be handled?

Should the reading of R1 rate be done per spectrum, matched to the spectrometer frequency and spin-lock offset, and then the values are averaged for those spectra having the same spectrometer frequency and spin-lock offset?

Best
Troels Emtekær Linnet


Related Messages